1 kW Electronic Load Details
Note: This unit was manufactured from 1998 through 2007 and is no longer in production.
Please refer to the Innovator Series Electronic Loads for current products.
Information is provided for reference only
|
Product Overview
|
+Self-contained air cooled system +Fully user programmable operation +Constant current, constant voltage, or constant power modes of operation +Automatic current ramp operation with user specified starting load, ending load, and ramp rate +Pulsed current operation: adjustable amplitude, duty cycle, and frequency +Externally located standard shunt for convenient access +Interchangeable shunts for precise operation over a wide dynamic range +20V and 55V models available +System wide control of all FiderisTM FCTS equipment using FCPowerTM Software |
|
Description
|
This patented design has been optimized to provide ultra-low impedance operation for fuel cell testing. The system is designed to provide maximum power with minimum resistance. This system is fully programmable and has a solid-state load. In addition, load sharing technology distributes currents among transistors, eliminating component overheating and improving reliability. |
|
Specifications
|
|
Loadbank
|
|
|
Type: |
MOSFET variable resistance loadbank |
|
|
Current Measurement: |
Exchangeable shunt (1 user-specified shunt provided, additional shunts available) |
|
|
Maximum Current Rating: |
1000 Amps |
|
|
Maximum Voltage Rating: |
20 Volts (55 Volts optional) |
|
|
Maximum Power Dissipation: |
1 kW standard |
|
|
Nominal Short Circuit Resistance: |
< 2 mohms |
|
|
Load Capacitance: |
Less than 25 pF |
|
|
Corrosion Protection: |
All critical components gold-plated |
|
Back Panel Connections
|
|
|
Power Inlet: |
IEC-320 (mating cord provided) |
|
|
Load Connections: |
Bolt down terminals standard |
|
Electrical Requirements
|
|
|
Voltage: |
120 VAC (208/220/240 optional) |
|
|
Frequency: |
60 Hz (50 Hz optional) |
|
|
Circuit Rating: |
3 Amps @ 120V |
|
Installation Requirements
|
|
|
Clearance for Operation: |
Back: 8", Sides: 0", Top: 0" |
|
Physical Characteristics
|
|
|
Dimensions: |
8" high, 18.5" wide, 21" deep |
|
|
System Weight: |
55 lbs. |
|
|
Operating Environment: |
40 to 100 Fahrenheit, 85% RH |
|
|
Cabinet Rating: |
Indoor |
|
|
Software
|
Fideris
FCPower
TM
software. |
|
Photographs
|
|
|
|
|
|
News:
TesSol, Inc, Announces Their Innovator Series Equipment Platform
. Superior design and testing flexibility.
|
|
|
|
|
|
|
|
|
|
|
|
| Copyright ©2008 TesSol, Incorporated |
|
|
home
| contact
| support
| sales
| faqs
products
| company information
| site map
Tes Sol, Inc. | 1315 SE Grace Ave, Ste 130 | Battle Ground, WA 98604 USA
Phone: (360) 687-1393, Fax: (360) 687-8251
Need
help? Contact
our webmaster.