Bubble Humidifier Details


View Brochure
high resolution (661kB PDF)
Product FAQs
[Product Overview] [Description] [Specifications] [Software]
Product Overview +Wide dynamic range of gas flows from 0 to 10 slpm
+Humidifier can be operated as a stand alone unit, with Fideris FCPowerTM control software, or as part of the Fideris Hydrogen Test KitTM
+Available with one or two gas humidification channels
+Independent temperature control of fuel and oxidant baths
+Front panel display of temperatures
+500 Watts per channel heating capacity
+Automatic water refill
+May be connected directly to house deionized water supply (water input pressure 10 psi greater than gas pressure required)
+Optional internal water pressurization pump allows use from ambient pressure water reservoir
+Swagelok valved front panel quick connects make connecting or disconnecting dry and easy
+Water handling system designed for use with deionized water
+System designed for reactant gas pressures up to 50 psig
+Reactant gas wetted materials limited to 316 stainless steel, Fluorocarbon FKM, silicon, and PTFE
+Designed for ease of diagnostics and maintenance
+Independent temperature control of optional heated gas transfer lines
+Hardwired Emergency Stop interlock inputs included for safety
+Hardwired Emergency Stop interlock outputs available
+Optional toxic and combustible gas detectors

Description
Specifications
Humidifier Specifications
Humidifiers: Evaporators equipped with temperature controlled 500 Watt heaters and gas diffusers
Evaporator Outlet Pressure: 50 psi maximum
Reactant Gas Flow Range: 0-10 slpm
Range of Temperature Control: Ambient to 115C
Over Pressure Relief Valve: 75 psi
Over Temperature Cutout: 120C
Gas Transfer Line Output
Voltage/Frequency: Same as system input
Max Current: 2 Amps each receptacle
Thermocouple Input: Mini Type K
Hardwired Emergency Interlocks
Inputs: Four contact closure inputs
Optional Outputs: Three contact closure outputs
Three 24VDC switched voltage outputs
Front Panel Connections
Fuel Supply and Return: 1/4" Swagelok quick-connects
Oxidant Supply and Return: 1/4" Swagelok quick-connects
Deionized Water Input: Quick-connect
Reservoir Level Switch:
(provided w/ optional reservoir)
LEMO
Fuel and Oxidant Humidifier Drain Lines: 1/4" Swagelok quick-connects
Optional Transfer Line Temperature Inputs: Mini-type K
Optional Transfer Line Switched Receptacles: IEC-320 outputs operating at same line voltage as the unit
Back Panel Connections
Power Inlet: IEC-320 (mating cord provided)
Overpressure Relief Valve: 1/4" Swagelok
Data Connection: RS-485 DB9
Hardwired Emergency Stop Interlock Inputs: Two 4 pin AMP circular connectors
Hardwired Emergency Stop Interlock Outputs (optional): Two 4 pin AMP circular connectors
Electrical Requirements
Voltage: 115 or 230 VAC as specified
Frequency: 50/60 Hz
Current Rating: 13 Amps at 115 VAC
10Amps at 230 VAC
Physical Characteristics
Dimensions: 27.5 D x 21.5 H x 10 W
System Weight: 85 to 95 pounds depending on options
Water Reservoir Capacity: 4.7 liters (optional)
Operating Environment: 40 to 100 Fahrenheit; 85% RH
Cabinet Rating: Indoor
Installation Requirements
Clearance for Operation: Back: 6, RH: 2, LH: 4, Top: 0
Software Fideris FCPower TM software, National Instruments LabView.
Photographs
News:
TesSol, Inc, Announces Their Innovator Series Equipment Platform .    Superior design and testing flexibility.
News

home | contact | support | sales | faqs
products | company information | site map

Tes Sol, Inc. | 1315 SE Grace Ave, Ste 130 | Battle Ground, WA 98604 USA
Phone: (360) 687-1393, Fax: (360) 687-8251

Need help? Contact our webmaster.