Product Overview +Independent mass flow control of up to four fuel gases and two oxidant gases (one fuel gas and two oxidant gases standard)
+Full Automation Capabilities with Fideris FCPowerTM Software
+Programming with Microsoft's VBScript is far superior to any other automation language
+Control gas composition
+Easily exchanged mass flow controllers
+Individual pressure regulators on each input gas line allows the use of dedicated gas cylinders or multi-user, high pressure gas lines
+Pressure gauges for regulated gas pressure monitoring and adjustment
+Check valves on inlet gas lines to prevent gas backflow into gas source lines
+All gas lines are made of seamless 316 stainless steel tubing
+3/8 tubing for high gas flows
+Backpanel access to static pressure ports allows user configuration of pressure monitoring points
+Fail safe purge of fuel system
+Panic button for emergency shutdown
+Ambient combustible and toxic gas detectors
+System wide control of all Fideris FCTS equipment using FCPowerTM Software(FCPowerTM Software is included)
Description The FiderisTM Gas Mixing System provides one oxidant and the blending of five fuels for reactant gas delivery to your fuel cell system. To provide the most accurate flow control, the pressure of the reactant gases on the inlet and the outlet of the mass flow controllers are regulated using precision forward pressure and backpressure regulators. Individual reactant gas purge lines allows the purge gas to be tailored to each side of the fuel cell and individual control of the purge provides maximum flexibility to the researcher. Finally, to provide the ultimate in control and safety, all reactant gas lines are provided with high quality valves on the inlet of each mass flow controller. Reactant gas sampling ports allow the operator to connect the Gas Mixing System to additional equipment for gas analysis.

As is expected from Fideris Incorporated, safety is addressed in every aspect of the design of the Gas Mixing Unit. For each gas channel an inlet pressure switch and mass flow controller inlet pressure gauge show the system and indicate to the control software that the sufficient reactant gas input is available. Pressure gauges on the mass flow controller discharge and on the exit manifolds provide a familiar and intuitive display of the status of the system to the operator. Non-adjustable, reactant gas overpressure switches provide for hardware shutdown should the reactant gas exit pressure exceed a specified limit. Additional hardware stops include a front panel emergency stop button, software trigger of the hardware emergency stop, and two external emergency stop inputs available to the user through the back panel. While both external emergency stop inputs can be used as desired, one is intended to interlock with FiderisTM Tail Gas Module to stop the flow of reactant gas to the fuel cell should the temperature of the tail gas exceed a user defined limit. Also available on the back panel is an electrically isolated hardware emergency stop output for connection to additional equipment such as a monitoring station, alarm, etc. Optional internal ambient gas detectors for combustible gas and/or toxic gas are available to provide early detection of hazardous conditions.
Specifications
Gas Handling
Reactant Gas Control: Individual mass flow control for up to four fuel gases and two oxidant gases
Range of Gas Flow: 1-10 lpm or 0-100 lpm standard
Maximum Inlet Gas Pressure: 250 psi
Back Panel Connections
Power Inlet: IEC-320 (mating cord provided)
Gas Inlets: 1/4" Swagelok for oxidant, fuel, and purge gases
Gas Outlets: 3/8" Swagelok
Data Connection: RS485 DB9
Electrical Requirements
Voltage: 120 VAC (208/220/240 optional)
Frequency: 60 Hz (50 Hz optional)
Current: 6 Amps @ 120V
Installation Requirements
Clearance for Operation: Back: 6, RH: 2, LH: 2, Top: 0
Physical Characteristics
Dimensions: 12.75" high; 21.75" wide; 25.65" deep
System Weight: 75 lbs.
Operating Environment: 40 to 100 Fahrenheit, 85% RH
Cabinet Rating: Indoor
Software Fideris FCPower TM software.
Photographs
News:
Fideris Brand Testing Equipment is Now Available Through the Fuel Cell Store Website .    The Fuel Cell Store is now a distributor for the Fideris Brand of testing and laboratory automation equipment.  This Partnership is the first to offer the industry component selection with on-line pricing, allowing customers to tailor their equipment to match their requirements; creating a plug-and-play, turn-key fuel cell test system.
News

home | contact | support | sales | faqs
products | company information | site map

Tes Sol, Inc. | 1315 SE Grace Ave, Ste 130 | Battle Ground, WA 98604 USA
Phone: (360) 687-1393, Fax: (360) 687-8251

Need help? Contact our webmaster.