Location: Home / Product H1212 |
|
|
H0606/H1212/H2412 Humidifier Details
 |

|
|
| [Product Overview] [Description] [Specifications] [Software] [Photographs] |
| Product Overview |
+Reactant gases heated only as necessary to entrain the desired quantity of water vapor +Water transfer rates of up to 24 mL/min +Easily filled water reservoir and pressurized water inlet +Water handling system is compatible with deionized water +System compatible with fuel cell backpressures up to 50 psig +Reactant gas handling systems are made of 316 stainless steel +Designed for ease of diagnostics and maintenance +Toxic gas detectors optional +Provides temperature control for external heaters such as gas transfer lines (transfer lines not included) +System wide control of all FiderisTM FCTS equipment using FCPowerTM Software (FCPowerTM Software is included) |
| Description |
The FiderisTM gas humidification unit is a component of the Fuel Cell Test System that has been developed to address all aspects of fuel cell research, testing, and utilization. In contrast to bubble type, or membrane based humidifiers, FiderisTM evaporator based humidifier allows a user-defined amount of water vapor to be entrained in the reactant gas streams. Overheating of the gas stream is eliminated since the gas is only heated to the temperature required to take up the specified amount of water. This technique provides the user with the ability to deliver any amount of water to the fuel cell, up to the maximum carrying capacity that may be entrained in the reactant gas stream. |
| Specifications |
| Humidification |
|
| Humidifiers:
| Flash evaporation directly into each reactant gas stream providing 100% RH
|
| Back Panel Connections |
|
| Power Inlet:
| NEMA L6-30P Twist Lock plug with 10' cord
|
|
| Gas Inlets:
| 3/8" Swagelok for oxidant, fuel, and purge gases
|
|
| Gas Outlets:
| 3/8" Swagelok
|
|
| Data Connection:
| RS485 DB9
|
|
| Thermocouples:
| Mini Type K
|
|
| External Heater:
| Same as system input voltage
|
|
| Hot Gas Overpressure Release:
| 1/4 MNPT
|
| Electrical Requirements |
|
| Voltage:
| 208 VAC (220/240 optional)
|
|
| Frequency:
| 60 Hz (50 Hz optional)
|
|
| Current:
| 22 Amps @ 120V
|
|
| Circuit Rating:
| 30 Amps @ 120V
|
| External Heater Control Output |
|
| Voltage/Frequency:
| Same as system input
|
|
| Max Current:
| 3.5 Amps each
|
|
| Thermocouple Input:
| Mini Type K
|
| Physical Characteristics |
|
| Dimensions:
| 20 D x 18.5 H x 7.56 W
|
|
| System Weight:
| 55 pounds without water
|
|
| Water Reservoir Capacity:
| 2 gallons
|
|
| Operating Environment:
| 40 to 100 Fahrenheit, 85% RH
|
|
| Cabinet Rating:
| Indoor
|
| Operating Parameters |
|
| Max. Gas Backpressure:
| 50 psig
|
|
| Max. Water Delivery Rate Fuel Gas: Oxidant Gas:
|
12 mL/min 24 mL/min
|
| Installation Requirements |
|
| Clearance for Operation:
| Back: 6, Front: 8, RH: 6, LH: 6, Top: 0 | |
| Software |
Fideris FCPowerTM software. |
| Photographs | | |
|
|
|
News:
TesSol, Inc, Announces Their Innovator Series Equipment Platform
. Superior design and testing flexibility.
|
|
|
|
|
|
|
|
|
|
|